#1 |
|
Problem |
No Phaco Power. |
Solution |
- Check that the console is set for Phaco and the desired power level is set.
- Check that the Phaco plug is secure in the connector in the front panel.
- Check for broken wires near the plug and rear end of hand piece.
- Check that Phaco tip is tight.
- Check that cocoon is loose within the hand piece.
- Make sure that the Open Infusion button is not open during Phaco.
- Make that the Phaco power light is on when the surgeon goes in to the third position of the foot switch.
|
#2 |
|
Problem |
Clouding in Anterior Chamber. |
Solution |
- Increase flow rate.
|
#3 |
|
Problem |
Incision Heating. |
Solution |
- Make sure size of the incision is correct for phaco tip.
- Replace Cobra Sleeve, internal ribs may be worn.
- Before starting ultrasound make sure you have flow and tip is not clogged with viscoelastic.
- Make sure that the phaco tip is in the center of the incision.
|