#1, #2, #3 |
|
Problem |
Valves don't close |
Solution |
- Clean valve using a syringe with warm distilled or sterile water, while simultaneously working the foot pedal. Inject a drop of lubricant into the plunger channel while working the foot pedal.
- Power off. Valve #3 can be manually moved by fitting your left index finger under the cover at a 45° angle, feeling a circular flat disc and physically working the valve. This is the most common valve to get stuck, since it is on the bottom and gets hit with the most BSS.
- Make sure, that lines are properly placed in the valves.
|
#4 |
|
Problem |
Pump not running properly |
Solution |
- Open the pump lock. Pull exiting end of tubing to give some tension and then close the lock with the other hand.
- Is pump tubing cramped? Rearrange it or change.
- Are rollers free to turn? If not, lubricate white plastic rollers with lubricant.
- Is it SDC pump tubing? Use SDC pump tubing only.
|
#5 |
|
Problem |
No suction at all |
Solution |
- Pump not locked? Lock it.
- Pump doesn’t turn? Open to check, than close.
- Shaft turns, but rollers don’t? Tighten the rollers to shaft.
- Notches on pump tubing? Either: massage pump tubing; pull pump tubing; or change pump tubing.
- Aspiration plug not plugged in? Check the O-rings and plug it in.
|
#6 |
|
Problem |
Clog in Aspiration Path |
Solution |
- Check Pressure Transducer. Unplug the aspiration line and read Actual Vacuum should be close to zero.
- Check reverse flow. Step on foot pedal to activate reflux switch.
- Test: Set flow rate to 12cc/min. Actual Vacuum should be below 100 mmHg.
If not, there is a clog, disconnect instrument from asp line, repeat test.
If OK, instrument is clogged. If not, aspiration line is clogged.
- Clean instrument or line to clear clogs, or change them.
|
#7 |
|
Problem |
Vacuum is not sufficient |
Solution |
- Is aspiration line placed properly in bottom valve?
- Does aspiration valve close completely? Try to pinch with fingers in step #7.
- Air bubbles in aspiration line? Is priming line dipped in container with water?
- Is pump running freely? Inspect pump tubing. Pull exiting end of
pump tubing. Change to new one, if worn.
- Check O-rings in aspiration plug.
- Does pressure transducer show zero when open?
- Test: Pinch aspiration line and set the flow rate to 20cc/min. Vacuum should easily reach preset 500 mmHg limit.
|
#9 |
|
Problem |
ultrasonic Hand Piece |
Solution |
- Tip is loose? Tip needs to be tightened using tip protector wrench.
- Have tips on all hand pieces been changed lately? Has anything been done to all the hand pieces recently?
- Inspect shell and rear connector
- Change Cocoon
- Cool hand piece down, try again.
- Change the hand piece. If problem is occurring with other hand pieces,
it may be the Ocusystem problem.
- Related: Metallic sound audible? Change O-rings.
|