NOTE: Pump Tubing should be replaced every 50 to 100 cases. |
#1 |
|
Problem |
Does not rotate. |
Solution |
- Check the pump tubing – should have talc on it and no kinks.
- Check rollers wheel that is flush with body and tight to the shaft.
- Open pump. Press the foot pedal check its indicator that it goes to Aspiration position. If motor does not run now, it is defective.
- Run motor with full speed. Grab the knob to stop motor, then release it. Must restart, otherwise it is too weak.
|
#2 |
|
Problem |
Noisy. |
Solution |
- Apply tiny drops of WD-40 between the rollers and the wall. On Manual mode ART Secondary Flow may equal 0. Turn it off.
- Apply tiny drops of WD-40 between the rollers and the wall. Check if pump touches the cover, if so - relocate the cover.
|
#2 |
|
Problem |
Bubbles. |
Solution |
- Was the set-up done properly? Re-prime.
- Inspect the cocoon – replace if worn out.
- Inspect hand-piece O-rings and rear connector.
|